A transportable x-ray fluorescence analysis system for assay of plutonium and uranium solutions in gloveboxes

Conference ·
OSTI ID:6683326
;  [1]; ; ;  [2];  [3]
  1. Lawrence Livermore National Lab., CA (USA)
  2. Commission of the European Communities, Luxembourg (Luxembourg). Safeguards Directorate
  3. Commission of the European Communities, Karlsruhe (Germany, F.R.). European Inst. for Transuranium Elements
It is becoming increasingly more difficult and expensive for safeguards authorities to transport solution samples from inspected facilities to laboratories for analysis. We have developed a transportable x-ray fluorescence analysis (XRFA) instrument that can assay plutonium and uranium solutions in glove boxes without any modifications to the glove box. This instrument will reduce the need to transport plutonium solution samples for analysis, thus saving time and reducing costs in inspection efforts. The system utilizes a portable high-purity germanium detector, transportable pulse-processing electronics, and a portable IBM-compatible computer. The measurement head of our instrument is placed inside a glove-box glove. Inside the glove box a hollow cylinder is held in place by the measurement head with the glove partially surrounding it, then solutions contained in 15-mL polystyrene test tubes are inserted into the cylinder for assay by XRFA. The radioisotope {sup 57}Co is used to excite the uranium and plutonium K-series x rays, and a combined {sup 57}Co and{sup 153}Gd source is used for transmission measurements to correct the observed x-ray intensity for sample self-attenuation effects. We have tested and evaluated the instrument at the European Institute for Transuranium Elements and demonstrated that 1% accuracies are achievable using eight uranium and seven plutonium reference solutions ranging in concentration from 50 to 411 g/L and 50 to 303 g/L, respectively. 3 refs., 2 figs., 2 tabs.
Research Organization:
Lawrence Livermore National Lab., CA (USA)
Sponsoring Organization:
DOE/DP
DOE Contract Number:
W-7405-ENG-48
OSTI ID:
6683326
Report Number(s):
UCRL-102705; CONF-9007106--55; ON: DE90015206
Country of Publication:
United States
Language:
English

Similar Records

Transmission-corrected x-ray fluorescence analysis of uranium and plutonium solutions using a dual transmission source
Conference · Wed Dec 31 23:00:00 EST 1986 · Transactions of the American Nuclear Society; (USA) · OSTI ID:5693148

A new measurement system for IAEA inspectors: the compact K-Edge densitometer
Conference · Sun Jul 01 00:00:00 EDT 1984 · Nucl. Mater. Manage.; (United States) · OSTI ID:6152836

Transmission-corrected x-ray fluorescence analysis of uranium and plutonium solutions using a dual transmission source
Conference · Mon Nov 23 23:00:00 EST 1987 · OSTI ID:5539056