Improved performance from modifications to the multidetector SPECT brain scanner

Journal Article · · J. Nucl. Med.; (United States)
OSTI ID:6681336

A multidetector single photon emission computerized tomographic brain scanner was modified to improve the angular sampling. The detector plate was rotated such that 12, 24, or 36 angular projections could be acquired. Phantom experiments demonstrated that the angular aliasing artifacts seen in images obtained with 12 detectors were eliminated with 36 effective detectors. In addition, the reconstructed image noise in a uniform source was decreased by a factor of 1.7 by the use of 36 instead of 12 angular projections, as predicted by computer simulation.

Research Organization:
Harvard Medical School, Boston, MA
OSTI ID:
6681336
Journal Information:
J. Nucl. Med.; (United States), Journal Name: J. Nucl. Med.; (United States) Vol. 25:6; ISSN JNMEA
Country of Publication:
United States
Language:
English