Dead time error with iodine-123 thyroid uptake measurements

Journal Article · · J. Nucl. Med. Technol.; (United States)
OSTI ID:6643388
Thyroid uptake measurements performed using iodine-123 (/sup 123/I) are subject to a systemic error due to dead time loss when the standard is derived based on a decay factor. Higher dead time loss with /sup 123/I is due to using more activity along with its better detection efficiency, higher count rate, and shorter half-life compared to that of /sup 131/I. The degree of error due to dead time loss may range from insignificant to extreme, depending on a number of factors. Degree of error is related to the activity administered, the window settings, the age and energy resolution of the instrumentation, and the elapsed time between standard and patient counts.