Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Attenuation maps for SPECT determined using cone beam transmission computed tomography

Conference · · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (USA)
OSTI ID:6609198
; ; ;  [1]
  1. Dept. of Radiology, SUNY Health Science Center, Syracuse, NY (US)
This paper presents a new method for measuring non-uniform attenuation maps, using a cone beam geometry CT scan acquired on a standard rotating gamma camera normally used for SPECT imaging. The resulting map is intended for use in non-uniform attenuation compensation of SPECT images. The method was implemented using a light-weight point source holder attached to the camera. A cone beam collimator may be used on the gamma camera, but the cone beam CT scans may also be acquired without collimator. In either implementation, the advantages include very high efficiency and resolution limited not by the collimator but by the intrinsic camera resolution (about 4 mm). Several phantoms tested the spatial uniformity, noise, linearity as a function of attenuation coefficient, and spatial resolution. Good quality attenuation maps were obtained, at least for the central slices where no truncation was present.
OSTI ID:
6609198
Report Number(s):
CONF-900143--
Conference Information:
Journal Name: IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (USA) Journal Volume: 37:2
Country of Publication:
United States
Language:
English