Half-cone beam collimation for triple-camera SPECT systems

Journal Article · · Journal of Nuclear Medicine
OSTI ID:394093
; ;  [1]
  1. North Carolina State Univ., Raleigh, NC (United States)

Cone-beam collimators provide increased sensitivity at similar resolution compared to other collimators. The use of cone-beam collimators for brain imaging with triple-camera SPECT systems, however, results in truncation of the base of the brain because of clearance of the shoulders. A half-cone beam collimator does not have the problem of truncation. The objective of this study was to compare the performance characteristics of half-cone beam with parallel-beam and fan-beam collimators with similar resolution characteristics for SPECT imaging of the brain. A half-cone beam collimator with the focal point located towards the base of the brain was built for a triple-camera SPECT system. Spatial resolutions and sensitivities of three collimators were measured. When 10-cm from the collimator surface, the planar spatial resolutions FWHM in mm (point source sensitivities in cps-MBq) for half-cone beam, fan-beam and parallel-beam collimators were 5.2 (85.6), 5.1 (55.6) and 5.9 (39.7), respectively. Image quality was evaluated using a three-dimensional Hoffman brain phantom and patient data. The deeper gray matter were more clearly visualized in the half-cone beam scans. Half-cone beam collimation provides higher sensitivity and offers the potential for improved brain imaging compared with parallel-beam and fan-beam collimation when used with a triple-camera SPECT system. 23 refs., 9 figs., 1 tab.

DOE Contract Number:
FG02-96ER62150
OSTI ID:
394093
Journal Information:
Journal of Nuclear Medicine, Journal Name: Journal of Nuclear Medicine Journal Issue: 3 Vol. 37; ISSN JNMEAQ; ISSN 0161-5505
Country of Publication:
United States
Language:
English

Similar Records

A combined half-cone beam and parallel hole collimation system for SPECT brain imaging
Journal Article · Mon Jun 01 00:00:00 EDT 1998 · IEEE Transactions on Nuclear Science · OSTI ID:649401

Collimator design for a multipinhole brain SPECT insert for MRI
Journal Article · Sat Nov 14 23:00:00 EST 2015 · Medical Physics · OSTI ID:22482411

Brain SPECT with short focal-length cone-beam collimation
Journal Article · Fri Jul 15 00:00:00 EDT 2005 · Medical Physics · OSTI ID:20726085