Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

A combined half-cone beam and parallel hole collimation system for SPECT brain imaging

Journal Article · · IEEE Transactions on Nuclear Science
DOI:https://doi.org/10.1109/23.682006· OSTI ID:649401
; ;  [1]
  1. Duke Univ. Medical Center, Durham, NC (United States). Dept. of Radiology

The research group has recently examined new types of collimator designs with the goal of improving sensitivity and lesion detection for head imaging. One of these collimator designs is a half-cone beam collimator. However, the sensitivity is reduced as objects are removed from the focal line and it does not satisfy Tuy`s sufficiency condition. Parallel hole collimation does not have this problem with axial blurring, however, values for sensitivity are not as large as for half-cone collimators. A potential configuration is to use one parallel hole collimator in conjunction with two half-cone beam collimators on a triple camera single photon emission computed tomography (SPECT) system. This might produce results that represent the best of both collimation systems. The authors acquired projection data with a Defrise disk phantom filled with Tc-99m. A half-cone beam collimator with a focal length of 50 cm was placed on one head of a triple camera SPECT system. A low energy super high resolution (LESR) parallel hole collimator was placed on a second head. Different projection data sets were combined to model acquisition in the three headed gamma camera with three half-cone beam collimators, two half-cone beam and one parallel beam collimators, one half-cone beam and two parallel beam collimators and three parallel beam collimators. Image reconstruction used a modified maximum likelihood maximization-expectation (ML-EM).

Sponsoring Organization:
National Cancer Inst., Bethesda, MD (United States); USDOE, Washington, DC (United States)
DOE Contract Number:
FG02-96ER62150
OSTI ID:
649401
Report Number(s):
CONF-971147--
Journal Information:
IEEE Transactions on Nuclear Science, Journal Name: IEEE Transactions on Nuclear Science Journal Issue: 3Pt2 Vol. 45; ISSN 0018-9499; ISSN IETNAE
Country of Publication:
United States
Language:
English

Similar Records

Half-cone beam collimation for triple-camera SPECT systems
Journal Article · Thu Feb 29 23:00:00 EST 1996 · Journal of Nuclear Medicine · OSTI ID:394093

Improved image quality for asymmetric double-focal cone-beam SPECT
Journal Article · Sun Aug 01 00:00:00 EDT 1993 · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States) · OSTI ID:6062230

Brain SPECT with short focal-length cone-beam collimation
Journal Article · Fri Jul 15 00:00:00 EDT 2005 · Medical Physics · OSTI ID:20726085