Six chemistry and three nondestructive assay (NDA) measurements have been made to provide information on the /sup 235/U content of twelve standards cut from a cylinder of uranium-aluminum alloy to be used in calibrating the /sup 252/Cf shuffler. These data have been evaluated to produce a set of uranium mass estimates and associated uncertainties for the standards by means of least squares techniques. Longitudinal fluctuation in uranium density is modeled, both by linear splines and by polynomials, and the two methods are compared. Also, a method is given for assessing the accuracy of the measurement uncertainties initially provided.
Research Organization:
Los Alamos National Lab., NM (USA)
DOE Contract Number:
W-7405-ENG-36
OSTI ID:
6554385
Report Number(s):
LA-8793-MS
Country of Publication:
United States
Language:
English

Similar Records

Test and evaluation results of the /sup 252/Cf shuffler at the Savannah River Plant
Technical Report · Sat Feb 28 23:00:00 EST 1981 · OSTI ID:6535082

Detectability limits and precision for shufflers
Technical Report · Wed Aug 01 00:00:00 EDT 1984 · OSTI ID:6528875

A shuffler for uranium billets
Conference · Mon Dec 31 23:00:00 EST 1990 · OSTI ID:5428952