Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

X-ray microcalorimeters with germanium resistance thermometers

Conference ·
OSTI ID:6539787
; ; ;  [1]; ; ; ; ;  [2]
  1. Lawrence Livermore National Lab., CA (USA)
  2. Lawrence Berkeley Lab., CA (USA)
We report on the current of our work on x-ray microcalorimeters for use as high resolution x-ray spectrometers. To maximize the x-ray collecting area and the signal to noise ratio, the total heat capacity of the device must be minimized. This is best achieved if the calorimeter is divided into two components, a thermal sensor and an x-ray absorber. The thermal sensor is a neutron transmutation doped (NTD) germanium resistor made as small as possible to minimize the heat capacity of the calorimeter. The thermistor can be attached to a thin x-ray absorber with large area and low heat capacity fabricated from superconducting materials such as niobium. We discuss results from our most recent studies of such superconducting absorbers and present the x-ray spectra obtained with these composite microcalorimeters at a temperature of 0.1 K. An energy resolution of 19 eV FWHM has been measured. 14 refs., 3 figs.
Research Organization:
Lawrence Livermore National Lab., CA (USA)
Sponsoring Organization:
DOE/DP
DOE Contract Number:
W-7405-ENG-48
OSTI ID:
6539787
Report Number(s):
UCRL-JC-104994; CONF-900756--49; ON: DE91000658
Country of Publication:
United States
Language:
English