The angular distribution of the avalanche around the anode wire in the gas proportional counter is determined by measuring the distribution of positive ions arriving on cathode strips surrounding the anode wire for each single event. The shape and width of the distribution depend on such factors as the gas gain, the anode diameter, the counting gas and the primary ionization density. Effects of these factors are studied systematically, and their importance for practical counter applications is discussed.