Data analysis for surface plate calibration

Technical Report ·
DOI:https://doi.org/10.2172/650156· OSTI ID:650156

A surface plate is used to establish a reference plane used for making precision dimensional measurements. Ideally, the surface should be perfectly flat. Since it is impossible to create a perfectly flat surface, surface plate are calibrated to determine how much they deviate from a perfectly flat surface. Once a surface plate is calibrated, the user can determine if the plate is flat enough for use in a particular application. The currently accepted method of data analysis for surface plate calibration, the Moody method, does not give optimum results because of the arbitrary way in which the reference plane is chosen. The exact solution, wherein the reference plane is located in the optimum position, has been worked out and is given in a step by step procedure along with all of the necessary equations. The amount of calculation involved in a typical example proved to be prohibitive even for a computer, so an alternate solution using a least squares criterion for locating the reference plane was developed. Results were comparable with those from Moody`s method with values being smaller or greater depending on the shape of the plate.

Research Organization:
AlliedSignal Federal Manufacturing and Technologies, Kansas City, MO (United States)
Sponsoring Organization:
USDOE, Washington, DC (United States)
DOE Contract Number:
AC04-76DP00613
OSTI ID:
650156
Report Number(s):
KCP--613-5986; ON: DE97054102
Country of Publication:
United States
Language:
English

Similar Records

Thickness effects on the plastic collapse of perforated plates with triangular penetration patterns
Conference · Tue Feb 29 23:00:00 EST 2000 · OSTI ID:770645

Interferometric study of natural-convection heat transfer from a vertical flat plate with transverse roughness elements
Thesis/Dissertation · Wed Dec 31 23:00:00 EST 1986 · OSTI ID:7255267

Calibration and equivalency analysis of image plate scanners
Journal Article · Fri Nov 14 23:00:00 EST 2014 · Review of Scientific Instruments · OSTI ID:22308646