Calibration and equivalency analysis of image plate scanners

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.4886390· OSTI ID:22308646
;  [1];  [2];  [3]
  1. Lawrence Livermore National Laboratory, 7000 East Avenue, Livermore, California 94550 (United States)
  2. Institute of Laser Engineering, Osaka University, Yamada-oka, 2-6, Suita, Osaka 565-0871 (Japan)
  3. Laboratory for Laser Energetics, University of Rochester, 250 East River Road, Rochester, New York 14623 (United States)

A universal procedure was developed to calibrate image plate scanners using radioisotope sources. Techniques to calibrate scanners and sources, as well as cross-calibrate scanner models, are described to convert image plate dosage into physical units. This allows for the direct comparison of quantitative data between any facility and scanner. An empirical relation was also derived to establish sensitivity response settings for arbitrary gain settings. In practice, these methods may be extended to any image plate scanning system.

OSTI ID:
22308646
Journal Information:
Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 11 Vol. 85; ISSN 0034-6748; ISSN RSINAK
Country of Publication:
United States
Language:
English

Similar Records

Sensitivity calibration of a Carestream HPX-1 image plate scanner
Technical Report · Fri Jan 31 23:00:00 EST 2025 · OSTI ID:2530803

On the system stability and calibration of the image plate/scanner system for plasma diagnosis at the National Ignition Facility
Journal Article · Wed Oct 17 00:00:00 EDT 2018 · Review of Scientific Instruments · OSTI ID:1479074

Cross-calibration of Fuji TR image plate and RAR 2492 x-ray film to determine the response of a DITABIS Super Micron image plate scanner
Journal Article · Mon Nov 14 23:00:00 EST 2016 · Review of Scientific Instruments · OSTI ID:22596625