Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Diesel exhaust/mine dust virtual impactor personal aerosol sampler. Design, calibration and field evaluation

Journal Article · · Aerosol Science and Technology; (United States)
; ;  [1]
  1. Univ. of Minnesota, Minneapolis, MN (United States)

A virtual impactor personal aerosol sampler (VIPAS) has been developed for use in diesel-equipped mines to collect size-separated samples of diesel exhaust and mine dust particles in the respirable size range. This sampler consists of three stages: a 10-mm cyclone, a 0.8-[mu]m outpoint virtual impactor and two filter cassettes. The large nonrespirable particles are removed in a 10-mm cyclone. The respirable particles from the cyclone pass through a 0.8-[mu]m cutpoint virtual impactor with the particles larger than and smaller than 0.8 [mu]m (mine dust and diesel exhaust particles, respectively) collected on two 37-mm filter cassettes. The sampling flow rate is 2 L/min so as to be compatible with the standard respirable dust sampler and personal sampling pumps. 22 refs., 10 figs.

OSTI ID:
6487632
Journal Information:
Aerosol Science and Technology; (United States), Journal Name: Aerosol Science and Technology; (United States) Vol. 22:2; ISSN 0278-6826; ISSN ASTYDQ
Country of Publication:
United States
Language:
English