Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Position-sensitive compressed-xenon electroluminescent gamma detector

Journal Article · · Instrum. Exp. Tech. (Engl. Transl.); (United States)
OSTI ID:6468821
A position-sensitive compressed-xenon electroluminescent gamma detector with a field of vision of 300 mm is described. A space resolution of 3.3 mm and an energy resolution of 4.3% are obtained for gamma radiation with an energy of 59.6 keV at a pressure of 0.8 MPa.
Research Organization:
All-Union Scientific-Research Institute of Medical Instrument Engineering, Moscow (USSR)
OSTI ID:
6468821
Journal Information:
Instrum. Exp. Tech. (Engl. Transl.); (United States), Journal Name: Instrum. Exp. Tech. (Engl. Transl.); (United States) Vol. 31:1; ISSN INETA
Country of Publication:
United States
Language:
English