Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Selective determination of chlorine dioxide using gas diffusion flow injection analysis

Journal Article · · Anal. Chem.; (United States)
DOI:https://doi.org/10.1021/ac00291a023· OSTI ID:6437802
An automated absorbance technique for the determination of aqueous chlorine dioxide has been developed by utilizing gas diffusion flow injection analysis. A gas diffusion membrane is used to separate the donor (sampling) stream from the acceptor (detecting) stream. The absorbance of chlorine dioxide is monitored at 359 nm. The first method uses distilled water as the acceptor stream and gives a detection limit of 0.25 mg/L chlorine dioxide. This system is over 550 times more selective for chlorine dioxide than chlorine. To further minimize chlorine interference, oxalic acid is used in the acceptor stream. The detection limit for this system is 0.45 mg/L chlorine dioxide. This second system is over 5400 times more selective for chlorine dioxide than chlorine. Both methods show excellent selectivity for chlorine dioxide over iron and manganese compounds, as well as other oxychlorinated compounds such as chlorite and perchlorate ions. 18 references, 7 figures, 3 tables.
Research Organization:
Miami Univ., Oxford, OH
OSTI ID:
6437802
Journal Information:
Anal. Chem.; (United States), Journal Name: Anal. Chem.; (United States) Vol. 57:14; ISSN ANCHA
Country of Publication:
United States
Language:
English