skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Sensitive bolometers using high-[ital T][sub [ital c]] superconducting thermometers for wavelengths 20--300 [mu]m

Journal Article · · Journal of Applied Physics; (United States)
DOI:https://doi.org/10.1063/1.354435· OSTI ID:6273995
;  [1]; ;  [2]
  1. Department of Physics, University of California and Materials Sciences Division, Lawrence Berkeley Laboratory, Berkeley, California 94720 (United States)
  2. Conductus Inc., Sunnyvale, California 94086 (United States)

Two types of sensitive high-[ital T][sub [ital c]] bolometer have been fabricated on 20-[mu]m-thick sapphire substrates with YBa[sub 2]Cu[sub 3]O[sub 7[minus][delta]] thin film transition edge thermometers. The sensitivity of both devices is superior to commercially available, room temperature pyroelectric detectors. A 1[times]1 mm bolometer with a radiation absorber of gold black smoke has a noise equivalent power NEP=2.4[times]10[sup [minus]11] W/Hz[sup 1/2] at 10 Hz and has useful sensitivity over wavelengths 20--100 [mu]m. A 3[times]3 mm bolometer with a Bi film radiation absorber has NEP=3.6[times]10[sup [minus]10] W/Hz[sup 1/2] at 10 Hz and has useful sensitivity over wavelengths 100--300 [mu]m. Water vapor transmittance spectra were measured to demonstrate the sensitivity of both devices.