Fabrication of an infrared bolometer with a high T sub c superconducting thermometer

Conference ·
OSTI ID:6184409
;  [1]; ;  [2]
  1. California Univ., Berkeley, CA (USA). Dept. of Physics Lawrence Berkeley Lab., CA (USA)
  2. Conductus, Inc., Sunnyvale, CA (USA)
A sensitive high {Tc} superconducting bolometer has been fabricated on a 20 {mu}m thick sapphire substrate with a YBCO thin film transition edge thermometer. Optical measurements with a He-Ne laser gave a noise equivalent power of 2.4{center dot}10{sup {minus}11} W/Hz{sup 1/2} at 10 Hz and a responsivity of 17 V/W in good agreement with electrical bolometer measurements. Gold black smoke was then deposited on the back side of the assembled bolometer as an absorber. Spectral measurements on a Fourier transform spectrometer show that the bolometer has useful sensitivity from visible wavelengths to beyond {approximately}100 {mu}m. This performance is clearly superior to that of a commercial room temperature pyroelectric detector. Some improvement appears possible. 10 refs., 5 figs.