skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Trace impurity analyzer

Conference ·
OSTI ID:6265043

The desirability for long-term reliability of large scale helium refrigerator systems used on superconducting accelerator magnets has necessitated detection of impurities to levels of a few ppM. An analyzer that measures trace impurity levels of condensable contaminants in concentrations of less than a ppM in 15 atm of He is described. The instrument makes use of the desorption temperature at an indicated pressure of the various impurities to determine the type of contaminant. The pressure rise at that temperature yields a measure of the contaminant level of the impurity. A LN/sub 2/ cryogenic charcoal trap is also employed to measure air impurities (nitrogen and oxygen) to obtain the full range of contaminant possibilities. The results of this detector which will be in use on the research and development helium refrigerator of the ISABELLE First-Cell is described.

Research Organization:
Brookhaven National Lab., Upton, NY (USA)
DOE Contract Number:
EY-76-C-02-0016
OSTI ID:
6265043
Report Number(s):
BNL-25754; CONF-790327-63; TRN: 79-011477
Resource Relation:
Conference: IEEE particle accelerator conference, San Francisco, CA, USA, 12 Mar 1979
Country of Publication:
United States
Language:
English