Title: Absolute detection efficiencies of microchannel plates for 0. 1--2. 3 keV electrons and 2. 1--4. 4 keV Mg ions

Journal Article · · Rev. Sci. Instrum.; (United States)
DOI:https://doi.org/10.1063/1.1138944· OSTI ID:6188601

The absolute detection efficiencies of detectors consisting of two microchannel plates (MCP) in a chevron arrangement, were experimentally determined for 0.1--2.3-keV electrons and 2.1--4.4-keV Mg ions. Both detectors tested included a grid with 92.5% transmission in front of the first MCP. For the measurements, the observed detector count rates were compared to the corresponding particle currents collected in a Faraday cup and measured with a vibrating reed electrometer. The electron detection efficiency of the MCP detector, including the grid, decreases from 0.82 at 0.1 keV to 0.65 at 2.3 keV for electrons incident normal to the surface. The Mg ion detection efficiency for the same arrangement, but with 43 incidence angle, increases from 0.49 at 2.1 keV to 0.81 at 4.4 keV.