Title: Calibration of bremsstrahlung prefogged Kodak RAR 2492 film

Journal Article · · Review of Scientific Instruments; (USA)
DOI:https://doi.org/10.1063/1.1141825· OSTI ID:6135290
; ; ; ; ;  [1];  [2]
  1. KMS Fusion, Inc., Ann Arbor, MI (USA)
  2. Sandia National Laboratory, Albuquerque, NM (USA)

High-energy background radiation from PBFA II at Sandia National Laboratory introduces uncertainty regarding the effect of background fogging on the sensitivity of the x-ray film at soft x-ray energies. We have performed a calibration to determine how the sensitivity of the Kodak RAR 2492 film is affected by high-energy background radiation. To simulate the background radiation the film was fogged to various densities using a 10 keV bremsstrahlung spectrum. The film was then exposed to soft x-ray emission lines of Al {ital K}{alpha} and Ti {ital K}{alpha} selected by Bragg reflection from an electron bombardment source. The intensity of the x-ray flux was continuously monitored with a Si(Li) detector to eliminate error due to drift of the x-ray source's intensity. A microdensitometer with matched objectives was used to find the specular density of the exposed film. The results of the calibration are presented in the form of {ital D} vs log {ital l} for the various densities of the bremmstrahlung prefog exposures.