Calibration of bremsstrahlung prefogged Kodak RAR 2492 film
- KMS Fusion, Inc., Ann Arbor, MI (USA)
- Sandia National Laboratory, Albuquerque, NM (USA)
High-energy background radiation from PBFA II at Sandia National Laboratory introduces uncertainty regarding the effect of background fogging on the sensitivity of the x-ray film at soft x-ray energies. We have performed a calibration to determine how the sensitivity of the Kodak RAR 2492 film is affected by high-energy background radiation. To simulate the background radiation the film was fogged to various densities using a 10 keV bremsstrahlung spectrum. The film was then exposed to soft x-ray emission lines of Al {ital K}{alpha} and Ti {ital K}{alpha} selected by Bragg reflection from an electron bombardment source. The intensity of the x-ray flux was continuously monitored with a Si(Li) detector to eliminate error due to drift of the x-ray source's intensity. A microdensitometer with matched objectives was used to find the specular density of the exposed film. The results of the calibration are presented in the form of {ital D} vs log {ital l} for the various densities of the bremmstrahlung prefog exposures.
- DOE Contract Number:
- AC03-87DP10560
- OSTI ID:
- 6135290
- Journal Information:
- Review of Scientific Instruments; (USA), Vol. 61:10; ISSN 0034-6748
- Country of Publication:
- United States
- Language:
- English
Similar Records
Low-energy x-ray response of photographic films. II. Experimental characterization
Absolute calibration of Kodak Biomax-MS film response to x rays in the 1.5- to 8-keV energy range
Related Subjects
PARTICLE BEAM FUSION ACCELERATOR
BACKGROUND RADIATION
PLASMA DIAGNOSTICS
X-RAY DETECTION
BREMSSTRAHLUNG
CALIBRATION
DATA ACQUISITION
PHOTOGRAPHIC FILM DETECTORS
ACCELERATORS
DETECTION
ELECTROMAGNETIC RADIATION
MEASURING INSTRUMENTS
RADIATION DETECTION
RADIATION DETECTORS
RADIATIONS
700102* - Fusion Energy- Plasma Research- Diagnostics