Design for a multiple target system for a medical cyclotron
Journal Article
·
· Med. Phys.; (United States)
A novel target system has been designed for a compact cyclotron. The system permits one of three or more targets to be selected remotely for bombardment, permits all target and chemical operations to be controlled remotely, causes no change in beam energy or beam current, occupies a limited space, and is simple to construct and maintain. The targets are mounted to a flexible bellows-multiple port manifold assembly, supported on a fan-shaped turntable. A motor moves the system to align the target to be bombarded with the main beam axis. Designs for two subassemblies are also presented. These are a retractable target chamber for water and a remotely controlled variable collimator.
- Research Organization:
- Biophysics Laboratory, Memorial Sloan-Kettering Cancer Center, New York, New York 10021
- OSTI ID:
- 6133099
- Journal Information:
- Med. Phys.; (United States), Journal Name: Med. Phys.; (United States) Vol. 8:5; ISSN MPHYA
- Country of Publication:
- United States
- Language:
- English
Similar Records
Design for a multiple target system for a medical cyclotron
Target device for making radionuclides in a cyclotron
Adjustable collimator
Journal Article
·
Wed Dec 31 23:00:00 EST 1980
· Med. Phys.; (United States)
·
OSTI ID:5610135
Target device for making radionuclides in a cyclotron
Journal Article
·
Fri May 01 00:00:00 EDT 1987
· Instrum. Exp. Tech. (Engl. Transl.); (United States)
·
OSTI ID:5843398
Adjustable collimator
Patent
·
Tue Jul 07 00:00:00 EDT 1981
·
OSTI ID:5857527