Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Transmission computed tomography data acquisition with a SPECT system

Journal Article · · J. Nucl. Med. Technol.; (United States)
OSTI ID:5994533

Phantom and animal transmission computed tomography (TCT) scans were performed with a camera-based single photon emission computed tomography (SPECT) system to determine system linearity as a function of object density, which is important in the accurate determination of attenuation coefficients for SPECT attenuation compensation. Results from phantoms showed promise in providing a linear relationship in measuring density while maintaining good image resolution. Animal images were essentially free of artifacts. Transmission computed tomography scans derived from a SPECT system appear to have the potential to provide data suitable for incorporation in an attenuation compensation algorithm at relatively low (calculated) radiation doses to the subjects.

Research Organization:
Duke Univ. Medical Center, Durham, NC
OSTI ID:
5994533
Journal Information:
J. Nucl. Med. Technol.; (United States), Journal Name: J. Nucl. Med. Technol.; (United States) Vol. 15:2; ISSN JNMTB
Country of Publication:
United States
Language:
English