On-line analysis of plutonium by energy dispersive x-ray fluorescence

Conference ·
OSTI ID:5977146

An on-line energy dispersive K x-ray fluorescence analysis (XRFA) system has been developed to measure the plutonium concentration of process solutions at the Savannah River Plant (SRP). This system has consistently produced accountability quality measurements (< 0.5%) on-line in approx.40 minutes of counting time. Comparisons to an independent coulometry method show agreement within sampling and analysis precision, typically 1%. 1 ref., 2 figs., 1 tab.

Research Organization:
Du Pont de Nemours (E.I.) and Co., Aiken, SC (USA). Savannah River Lab.; Lawrence Livermore National Lab., CA (USA)
DOE Contract Number:
W-7405-ENG-48; AC09-76SR00001
OSTI ID:
5977146
Report Number(s):
DP-MS-85-77; CONF-8510151-18; ON: DE86008901
Country of Publication:
United States
Language:
English