The authors have studied the reproducibility upon thermal cycling of several types of thermometers. A Josephson junction noise thermometer, a CMN thermometer, and an SRM 768 superconductive fixed-point device were very consistent, while a germanium and a carbon resistance thermometer showed significant irreproducibility.