Determination of the superconductive transition temperatures of Cd, Zn, and Al using a Josephson-junction noise thermometer. Final report

Technical Report ·
OSTI ID:5943771
In order to evaluate the accuracy of the EPT-76 temperature scale, the authors measured three fixed points defined on that scale (the superconductive transition temperatures of Cd, Zn, and Al) using a Josephson junction noise thermometer.