Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Evaluation of a continuous /sup 14/C, /sup 129/I, and /sup 85/Kr monitor under dissolver off-gas conditions

Technical Report ·
DOI:https://doi.org/10.2172/5920975· OSTI ID:5920975
A previously developed /sup 129/I, /sup 14/C, and /sup 85/Kr stack monitor was evaluated for use on dissolver off-gas streams. The effect of higher concentrations of H/sub 2/, Hg (a dissolution catalyst), and Cd (a common neutron poison) was measured in a series of laboratory tests. No significant effect on the sensitivity of the monitor was observed due to the presence of higher H/sub 2/ concentrations; but the variability did increase, probably due to condensation. Both Cd and Hg significantly reduced the sensitivity and increased the noise level of the monitor. Therefore, the /sup 129/I monitor, although suitable for stack monitoring, was considered unsuitable for dissolver off-gas streams containing high concentrations of Hg and Cd.
Research Organization:
Exxon Nuclear Idaho Co., Inc., Idaho Falls (USA)
DOE Contract Number:
AC07-79ID01675
OSTI ID:
5920975
Report Number(s):
ENICO-1135; ON: DE83017048
Country of Publication:
United States
Language:
English