Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Investigation of large format microchannel plate Z configurations

Conference · · IEEE Trans. Nucl. Sci.; (United States)
OSTI ID:5898181
The performance of triplet (Z) stacks of microchannel plates (MCP's) has been studied as a part of the instrument development for the Extreme Ultraviolet Explorer (EUVE) satellite mission. Relatively large MCP's with a 60 mm diameter and having an 80:1 channel length to diameter (L/D) ratio were used in several configurations. These have been used in the EUVE prototype imaging detector to provide > 512 X 512 pixels with low image distortion (< 1%). We have examined the gain and pulse height characteristics showing that both high gains (> 2 X 10/sup 7/) and tight output pulse height distributions (< 30% FWHM) may be achieved. Simple distribution techniques have also allowed low intrinsic background event rates (< .15 events cm/sup -2/ sec/sup -1/) to be obtained. Variation of the quantum efficiency of MCP's over the wavelength range 160 A to 1216 A has been investigated for a range of angles of incidence. The effect of temperature variations on MCP operating characteristics has also been evaluated.
Research Organization:
Univ. of California, Berkeley, CA
OSTI ID:
5898181
Report Number(s):
CONF-841007-
Conference Information:
Journal Name: IEEE Trans. Nucl. Sci.; (United States) Journal Volume: NS-32:1
Country of Publication:
United States
Language:
English