Calibration of densimeters for liquefied light hydrocarbons

Journal Article · · Oil Gas J.; (United States)
OSTI ID:5894953
Though the commercial densimeters tested all functioned in LNG, that is, all showed unambiguous variations in the parameter affected by density variations, they all required calibrations in liquids of known density and were found to have inadequate calibrations. Their calibration error was about ten times larger than their precision. The calibration error arose because the densimeter was sensitive to parameters not included in the calibration equation and also because the densimeters were never calibrated in LNG or methane for reasons of safety. Tests of calibration stability with time and use, performed on some of the commercial densimeters, indicated they had short term stability. The measurements of longer term stability were not done for some and inconclusive for those for which some tests were done.