Evaluation of the Beijing Instrument Factory GDB-44F photomultipler
Characteristics have been measured for the GDB-44F 2''-diameter photomultiplier. Some typical photomultiplier characteristics - such as gain, dark current, anode output current, transit and rise times as a function of voltage between anode and cathode, and photoelectron pulse-height spectrum, the quantum-efficiency and the single-electron transit time spread were measured. Descriptions of the measuring systems were also given.