Direct detection of vacuum ultraviolet radiation through an optical sampling orifice: determination of nonmetals in gaseous samples by inductively coupled plasma atomic emission spectroscopy

Journal Article · · Anal. Chem.; (United States)
DOI:https://doi.org/10.1021/ac00152a007· OSTI ID:5572909
A copper cone with a sampling orifice is attached to the entrance slit chamber or a monochromator and inserted directly into an inductively couples plasma (ICP) to sample vacuum ultraviolet (vacuum UV) radiation emitted by the plasma. A unique interface with low dead volume (<40 ..mu..L) is used for introduction of gaseous samples directly into the axial channel of the ICP. Injection of a gaseous mixture of compounds containing the elements Br, C, Cl, and S results in detection limits of 50, 30, 80 and 20 pg, respectively. These detection limits are by far the best obtained for these elements by ICP atomic emission spectroscopy (AES) and are comparable to those obtained by AES with He plasmas. Precision of the analyte emission intensities is 2% relative standard deviation or better.
Research Organization:
Iowa State Univ., Ames
DOE Contract Number:
W-7405-ENG-82
OSTI ID:
5572909
Journal Information:
Anal. Chem.; (United States), Journal Name: Anal. Chem.; (United States) Vol. 60:1; ISSN ANCHA
Country of Publication:
United States