The low-energy toroidal grating monochromator beamline at the synchrotron radiation source at Daresbury Laboratory

Journal Article · · Review of Scientific Instruments; (United States)
DOI:https://doi.org/10.1063/1.1143098· OSTI ID:5521115
 [1]; ;  [2]
  1. Department of Physics, University of Cambridge, Madingley Road, Cambridge, CB3 0HE (United Kingdom)
  2. SERC Daresbury Laboratory, Warrington, WA4 4AD (United Kingdom)
The bending magnet beam line 1.2 low-energy toroidal grating monochromator of the synchrotron radiation source (SRS) at Daresbury Laboratory, was designed{sup 1} to deliver moderately high fluxes ({similar to} 5 {times} 10{sup 11} photons s{sup {minus}1}), of linearly polarized, medium resolution ({similar to}0.2 eV) radiation in the energy range 5--85 eV. The colinear optical system utilizes platinum-coated silicon-carbide mirrors to focus the broad-band radiation emergent from the SRS at the entrance slits of the three grating monochromator. A single ellipsoidal mirror is then used to doubly focus the desired narrow-band radiation at the sample position. The colinear arrangement of the optical elements ensures that the radiation at the sample point is strongly horizontally plane polarized (estimated to be of order 90%). The entire system has been the subject of detailed analyses using the raytracing program SHADOW,{sup 2} and standard optical theory. These calculations have been compared with experimental determinations of photon flux outputs and resolution measurements.
OSTI ID:
5521115
Journal Information:
Review of Scientific Instruments; (United States), Journal Name: Review of Scientific Instruments; (United States) Vol. 63:1; ISSN RSINA; ISSN 0034-6748
Country of Publication:
United States
Language:
English

Similar Records

The high energy spherical grating monochromator minus A new source of soft x rays at Daresbury (abstract)
Journal Article · Tue Dec 31 23:00:00 EST 1991 · Review of Scientific Instruments; (United States) · OSTI ID:5569655

High resolution in the soft x-ray range from a toroidal grating monochromator
Journal Article · Wed Sep 01 00:00:00 EDT 1993 · Review of Scientific Instruments; (United States) · OSTI ID:6227684

A Soft X-Ray (300-1000 eV) Active Grating Monochromator Beamline at NSRRC
Journal Article · Thu Jan 18 23:00:00 EST 2007 · AIP Conference Proceedings · OSTI ID:21052579