Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Digital subtraction angiography system evaluation with phantoms

Conference · · Radiology; (United States)
OSTI ID:5482096

Advances in digital subtraction angiography imaging demonstrate the need for critical evaluation of the performance of digital subtraction equipment. The design of a phantom set for noninvasive assessment of the imaging quality of digital subtraction equipment is described; components include a remotely controlled transport system and individual patterns to evaluate the contrast and detail properties of the image intensifier, low-contrast sensitivity and resolution of the system, geometric distortion of image, linearity, mechanical and electronic stability of equipment, and effects of bone and bowel gas on iodine perception. The performance of an add-on digital radiographic system is presented, along with radiation exposure levels at the image intensifier for a range of radiographic techniques.

Research Organization:
Lahey Clinic Medical Center, Burlington, MA
OSTI ID:
5482096
Report Number(s):
CONF-851152-
Journal Information:
Radiology; (United States), Journal Name: Radiology; (United States) Vol. 155:3; ISSN RADLA
Country of Publication:
United States
Language:
English