Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Digital subtraction angiography systems: assessing performance with phantoms

Conference · · Radiology; (United States)
OSTI ID:5212039
A solid, acrylic phantom was developed to assess the iodine perceptibility of three digital subtraction angiography (DSA) systems, including one using a recursive filter technique. Three phantoms, designed to mimic flow of iodinated boluses through blood vessels, were used. Each contained four triangular boluses of different widths and of a specific iodine concentration. Motion, required to test recursively filtered DSA systems, was simulated by placing the phantoms on a motorized cart. The phantoms were also used to test flow curve generation in applicable systems. Intersystem differences in perceptibility were attributed to differences in system noise, system resolution, and dynamic range. Flow curves produced with the phantoms were similar to those obtained in patient studies.
Research Organization:
Massachusetts General Hospital, Boston
OSTI ID:
5212039
Report Number(s):
CONF-851152-
Conference Information:
Journal Name: Radiology; (United States) Journal Volume: 156:3
Country of Publication:
United States
Language:
English