A 14 MeV absolute neutron beam has been established at the NBS 3 MV positive-ion Van de Graaff Accelerator Laboratory. The neutron flux is absolutely determined by the measurement of the associated alpha particle rate in a silicon surface barrier detector positioned at 84 degrees with respect to the beam axis. The facility may be used to measure the absolute response of active neutron monitoring devices with high accuracy. The neutron background contributions may be eliminated in the calibration measurements by using the coincidence between the device and the solid state alpha detector. This method allows a neutron flux in a cone of 2 degree half angle to be placed at 90 degrees with respect to the beam axis.
Research Organization:
National Bureau of Standards, Washington, DC (USA)
OSTI ID:
5456226
Report Number(s):
PB-82-100181
Country of Publication:
United States
Language:
English

Similar Records

The NBS 14 MeV absolute neutron beam facility
Conference · Tue Mar 31 23:00:00 EST 1981 · IEEE Trans. Nucl. Sci.; (United States) · OSTI ID:5481021

NBS facilities for standardization of neutron dosimetry from 0. 001 to 14 MeV
Conference · Sat Oct 01 00:00:00 EDT 1977 · Natl. Bur. Stand. (U.S.), Spec. Publ.; (United States) · OSTI ID:5161379

Absolute measurement of the uranium-235 fission cross section from 0. 2 to 1. 2 MeV