Proton imaging for medical applications

Conference · · Radiology; (United States)
OSTI ID:5245846
In order to demonstrate the potential advantages of proton imaging for medical use, a 205-MeV proton beam was developed using the Argonne National Laboratory Booster I synchrotron. Data were taken using a narrow scanning beam and an electronic detector system. The proton images were compared to conventional radiographs, and demonstrate a significant dose reduction and an improved mass resolution over the radiographs. The images also show significant differences in the proton stopping power of different tissues, thus demonstrating considerable potential in soft-tissue imaging. The development of a proton tomographic scan system is also briefly discussed.
Research Organization:
Argonne National Lab., IL
OSTI ID:
5245846
Conference Information:
Journal Name: Radiology; (United States) Journal Volume: 135:2
Country of Publication:
United States
Language:
English