Compensated count rate circuit for portable Geiger-Mueller survey meter

Journal Article · · Rev. Sci. Instrum.; (United States)
DOI:https://doi.org/10.1063/1.1136359· OSTI ID:5159913
A count rate circuit is described that better compensates for counting losses due to detector dead time than previously used techniques in Geiger--Mueller survey meters. Bench tests show the new design response to be within 1% of the ideal response even up to counting losses of 50%.
Research Organization:
Oak Ridge National Laboratory, Oak Ridge, Tennessee 37830
DOE Contract Number:
W-7405-ENG-26
OSTI ID:
5159913
Journal Information:
Rev. Sci. Instrum.; (United States), Journal Name: Rev. Sci. Instrum.; (United States) Vol. 51:8; ISSN RSINA
Country of Publication:
United States
Language:
English