Large-entry-window detector consisting of a microchannel plate and a channel-type electron multiplier

Journal Article · · Instrum. Exp. Tech. (Engl. Transl.); (United States)
OSTI ID:5130863
The bas'c parameters of an assembly consisting of a microchannel plate and a channel-type electron multiplier have been measured. The counting rate limit is 10/sup 5/ sec/sup -1/ in this recording unit at a trigger threshold of 2.5 mV. The dependencies of both the counting efficiency and the amplitude distribution upon the input load, the supply voltage, and the displacement voltage are described.