Title: Elemental speciation by anion exchange and size exclusion chromatography with detection by inductively coupled plasma mass spectrometry with direct injection nebulization

Journal Article · · Analytical Chemistry (Washington); (United States)
DOI:https://doi.org/10.1021/ac00069a006· OSTI ID:5119411
;  [1]
  1. Ames Lab., IA (United States) Iowa State Univ., Ames, IA (United States)

A direct injection nebulizer (DIN) is used with packed microcolumns for anion chromatography (AC) and size exclusion chromatography (SEC). Two Se species (SeO[sub 3][sup 2[minus]] and SeO[sub 4][sup 2[minus]]) are separated by AC with detection limits of 15 pg of Se. The isotope ratio [sup 74]Se/[sup 78]Se can be measured with a relative standard deviation of 0.5% on 25 ng of each Se species during the actual chromatographic separation. Proteins in human serum are separated by SEC without sample pretreatment. The metals present in each molecular weight fraction are determined by ICP-MS with detection limits of 0.5-3 pg of metal. These absolute detection limits are 1-2 orders of magnitude better than those obtained with conventional nebulizers. 56 refs., 3 figs., 7 tabs.

DOE Contract Number:
W-7405-ENG-82
OSTI ID:
5119411
Journal Information:
Analytical Chemistry (Washington); (United States), Vol. 65:21; ISSN 0003-2700