Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Investigation on the low pressure limit of the Bayard--Alpert gauge

Journal Article · · J. Vac. Sci. Technol., A; (United States)
DOI:https://doi.org/10.1116/1.574063· OSTI ID:5032219
To extend the pressure measurement capability of a Bayard--Alpert (BA) gauge, several commercial BA gauges are modified. To study the position dependence of filament on gauge sensitivity, hair-pin filaments instead of commercial helical filaments were used. The characteristics of the modified gauge, which included the variation of the gauge sensitivity on the applied voltage and the position of the filament, are reported. To monitor the normal and the reverse x-ray photocurrents, the gauge was inserted into a fixed size tube of 304 LN stainless steel. The validity of the self-modulation was confirmed. A sensitivity of 60 Torr/sup -1/ and an x-ray limit of 2.7 x 10/sup -13/ Torr (N/sub 2/ equivalent) was obtained.
Research Organization:
Brookhaven National Laboratory, Upton, New York 11973
OSTI ID:
5032219
Journal Information:
J. Vac. Sci. Technol., A; (United States), Journal Name: J. Vac. Sci. Technol., A; (United States) Vol. 4:5; ISSN JVTAD
Country of Publication:
United States
Language:
English

Similar Records

Investigation on low pressure limit of Bayard-Alpert gauge
Conference · Fri Dec 31 23:00:00 EST 1982 · OSTI ID:5653528

Modified Bayard-Alpert Gauge with X-Ray Suppression
Journal Article · Mon Dec 31 23:00:00 EST 1962 · Review of Scientific Instruments · OSTI ID:4663139

Effect of magnetic field on the performance of Bayard--Alpert gauges
Journal Article · Sun Jan 31 23:00:00 EST 1982 · J. Vac. Sci. Technol.; (United States) · OSTI ID:5634621