Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Coincidence sum peak assay of /sup 125/I activity using solid-state detectors

Journal Article · · Med. Phys.; (United States)
OSTI ID:5012115
The use of solid-state detectors in the coincidence sum peak technique of /sup 125/I activity assay has been studied. The improved energy resolution of these detectors allows distinct resolution of the K..cap alpha.., K..beta.., and ..gamma.. peaks, and four sum peaks resulting from the coincident detection of the different X and ..gamma.. photons. Theoretical considerations yield three separate equations, each of which can be used, in conjunction with the experimental sum peak spectrum, to give the activity of the /sup 125/I sample in question. This method is suitable for calibration of /sup 125/I activities in the range of 0.01--0.5 ..mu..Ci (1 Ci=3.7 x 10/sup 10/ Bq) with precisions of 2%--3%.
Research Organization:
Division of Radiation Biophysics, Department of Radiation Medicine, Massachusetts General Hospital, Harvard Medical School, Boston, Massachusetts 02114
OSTI ID:
5012115
Journal Information:
Med. Phys.; (United States), Journal Name: Med. Phys.; (United States) Vol. 7:5; ISSN MPHYA
Country of Publication:
United States
Language:
English