Some problems associated with the 4$pi$$beta$--$gamma$ method (in French)

Technical Report ·
OSTI ID:4934048
4 pi BETA -- gamma counting are discussed: background, dead times, and finite resolving time of the coincidence circuit. The results of a Monte- Carlo simulation are presented: transmission factor for true coincidences, ratio of the random coincidence and true coincidence rates as a function of the delay between the BETA and gamma channels. (FR)