High rate 4. pi. beta. -. gamma. coincidence counting system

Conference ·
OSTI ID:6512349
A high count rate 4..pi.. ..beta..-..gamma.. coincidence counting system for the determination of absolute disintegration rates of short half-life radionuclides is described. With this system the dead time per pulse is minimized by not stretching any pulses beyond the width necessary to satisfy overlap coincidence requirements. The equations used to correct for the ..beta.., ..gamma.., and coincidence channel dead times and for accidental coincidences are presented but not rigorously developed. Experimental results are presented for a decaying source of /sup 56/Mn initially at 2 x 10/sup 6/ d/s and a set of /sup 60/Co sources of accurately known source strengths varying from 10/sup 3/ to 2 x 10/sup 6/ d/s. A check of the accidental coincidence equation for the case of two independent sources with varying source strengths is presented.