A METHOD FOR SUPPRESSING PHOTOMULTIPLIER NOISE
Technical Report
·
OSTI ID:4714509
The system was designed to take advantage of the difference in pulse length of slow-decay-time scintillators, such as CsI(Tl) and NaI(Tl), and the photomultiplier noise. The design has the advantage of simplicity and of using mostly standard electronic components, and is similar to the pulse shape discriminator system used to distinguish particle types. A typical spectrum obtained using the suppressor with a 8-kev x ray from Fe/sup 56/ on a CsI(Tl) crystal mounted on an RCA 6801A photomuitiplier is shown. (B.O.G.)
- Research Organization:
- California. Univ., Livermore. Lawrence Radiation Lab.
- DOE Contract Number:
- W-7405-ENG-48
- NSA Number:
- NSA-17-027657
- OSTI ID:
- 4714509
- Report Number(s):
- UCRL-6886(Rev.1)
- Country of Publication:
- United States
- Language:
- English