Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Quantitative analysis of hydrogen isotopes in the metal hydride of the neutron tube target

Technical Report ·
DOI:https://doi.org/10.2172/471443· OSTI ID:471443
We describe an experimental system for the quantitative analyses of hydrogen isotopes in metal hydride targets. The experimental system consists of a target desorption inlet system (TDIS) and a high-resolution mass spectrometer. The TDIS has a unique design that provides complete desorption of the gas and rapid, direct, and accurate measurements of pressure, volume, and temperature of the total gas desorbed from the target. The mass spectrometer measures partial pressures of the hydrogen isotopes from which the hydrogen isotopic ratios can be calculated. We present the operational parameters for optimum performance of the TDIS and results obtained from gas analyses of erbium hydride targets.
Research Organization:
Los Alamos National Lab., NM (United States)
Sponsoring Organization:
USDOE, Washington, DC (United States)
DOE Contract Number:
W-7405-ENG-36
OSTI ID:
471443
Report Number(s):
LA--13267-MS; ON: DE97005918
Country of Publication:
United States
Language:
English

Similar Records

Hydrogen isotope measurements for neutron tube targets
Conference · Fri Dec 31 23:00:00 EST 1976 · OSTI ID:6823157

Controlled Release of Hydrogen Isotopes from Hydride-Magnetic Nanomaterials
Journal Article · Wed Jan 29 19:00:00 EST 2020 · ACS Applied Materials and Interfaces · OSTI ID:1637908

Hydride compressor