GAMMA RAY SPECTROSCOPY USING A GERMANIUM LITHIUM-DRIFTED DIODE
Journal Article
·
· Nucl. Instr. Methods
Ge diodes with Li-drifted depletion layers are shown to be advantageous for gamma ray spectroscopy. The 122 and 136 kev lines of the Co/sup 57/ spectrum are readily resolved from the Compton background. The Cs/sup 137/ spectrum has a photoelectric peak about twice the height of the Compton background. It is pointed out that a Ge Li-drifted diode will require additional drifting after each use. (D.L.C.)
- Research Organization:
- RCA Victor Co. Ltd., Montreal
- NSA Number:
- NSA-17-025645
- OSTI ID:
- 4696273
- Journal Information:
- Nucl. Instr. Methods, Journal Name: Nucl. Instr. Methods Vol. Vol: 22
- Country of Publication:
- Country unknown/Code not available
- Language:
- English