S>The whole-body counter of Geneva was calibrated for Cs/sup 137/ measuremerts by using a Cs/sup 137/ standard solution and a Cs/sup 137/ standard solution, furnished by the IAEA. One energy band 150 kev wide from 575 to 726 kev was selected as the Cs/sup 137/ photopeak band. The measurements, made on six cases, yield a calibration factor of 7510 counts per minute per mu C of Cs/ sup 137/ in the body. The sensitivity is plus or minus 2%. (auth)
Research Organization:
Centre de Radioactivite Medicale, Geneva
NSA Number:
NSA-17-029208
OSTI ID:
4685657
Journal Information:
Helvetica Chimica Acta (Switzerland), Journal Name: Helvetica Chimica Acta (Switzerland) Vol. Vol: 46; ISSN HCACA
Country of Publication:
Country unknown/Code not available
Language:
French

Similar Records

CALIBRATION OF CAESIUM-137 CONTENT IN THE HUMAN BODY
Journal Article · Sat May 11 00:00:00 EDT 1963 · Nature · OSTI ID:4712765

ON THE CALIBRATION OF WHOLE BODY COUNTERS
Journal Article · Wed Feb 28 23:00:00 EST 1962 · Australian J. Appl. Sci. · OSTI ID:4833503

THE GENEVA COUNTER. A LARGE LIQUID SCINTILLATION COUNTER FOR MEN AND MESONS
Technical Report · Tue Jul 01 00:00:00 EDT 1958 · OSTI ID:4295232