OSTI ID:4333045
In the conventional procedure of isotopic analysis with the mass spectrometer, an observed ratio of mol ratios is corrected for an instrument phenomenon called memory Accurate analysis requires accurate knowledge of this correction factor. To demonstrate this requirement, a brief description of the single standard method is given Observed variations in this phenomenon within instruments and differences between instruments led to an investigation of a method which would reduce or eliminate the effect of memory. The essential features of an interpolative method are described, test results reported, and the direction of further study indicated. Comparative data are shown demonstrating the greater precision attained with the interpolative method. Error expressions for the single standard and interpolative methods also are presented (auth)
Research Organization:
Goodyear Atomic Corp., Portsmouth, Ohio
DOE Contract Number:
AT-(33-2)-1
NSA Number:
NSA-12-000808
OSTI ID:
4333045
Report Number(s):
GAT-171(Rev. 1)
Country of Publication:
United States
Language:
English

Similar Records

URANIUM HEXAFLUORIDE ISOTOPIC MEASUREMENTS USING AN INTERPOLATIVE METHOD
Technical Report · Tue Nov 29 23:00:00 EST 1960 · OSTI ID:4081249

PRECISION MASS SPECTROMETRIC DETERMINATIONS OF URANIUM ISOTOPIC COMPOSITION
Technical Report · Thu May 21 00:00:00 EDT 1959 · OSTI ID:4223214

A PROCEDURE FOR BLENDING INACCURATELY KNOWN BASES TO MATCH A U$sup 235$ ISOTOPIC STANDARD
Technical Report · Sun May 01 00:00:00 EDT 1960 · OSTI ID:4183460