PRECISION MASS SPECTROMETRIC DETERMINATIONS OF URANIUM ISOTOPIC COMPOSITION
Technical Report
·
OSTI ID:4223214
Mass spectrometers are subject to a variable phenomenon called "memory" which amounts to 1 to 2O% in the analysis of UF/sub 6/. A procedure using an interpolative method of standardization reduces the effect of memory to a negligible amount while at the same time compensating for or eliminating other sources of error. The use of a mass spectrometer for rapid and accurate determination of uranium isotopic abundance is described. (J.E.D.)
- Research Organization:
- Goodyear Atomic Corp., Portsmouth, Ohio
- DOE Contract Number:
- AT(33-2)-1
- NSA Number:
- NSA-14-003507
- OSTI ID:
- 4223214
- Report Number(s):
- GAT-T-664
- Resource Relation:
- Other Information: Presented at the American Society for Testing Materials, Committee e-14, Meeting on Mass Spectrometry, Los Angeles, Calif., May 21, 1959. Orig. Receipt Date: 31-DEC-60
- Country of Publication:
- United States
- Language:
- English
Similar Records
URANIUM HEXAFLUORIDE ISOTOPIC MEASUREMENTS USING AN INTERPOLATIVE METHOD
AN INTERPOLATIVE METHOD OF ASSAYING UF$sup 6$WITH THE ISOTOPIC MASS SPECTROMETER
Development of a portable mass spectrometric system for determination of isotopic composition of solid uranium samples using fluorine volatilization. Final report
Technical Report
·
Wed Nov 30 00:00:00 EST 1960
·
OSTI ID:4223214
AN INTERPOLATIVE METHOD OF ASSAYING UF$sup 6$WITH THE ISOTOPIC MASS SPECTROMETER
Technical Report
·
Sun Apr 01 00:00:00 EST 1956
·
OSTI ID:4223214
Development of a portable mass spectrometric system for determination of isotopic composition of solid uranium samples using fluorine volatilization. Final report
Technical Report
·
Fri Sep 30 00:00:00 EDT 1994
·
OSTI ID:4223214