Mercury cathode cell design for controlled potential coulometry
A comparison has been made of the characteristics of several mercury- cathode cells for use in controlled potential coulometry for the determination of uranium. The investigation showed that completely rigid control of the potential is not needed to achieve reliable titrations, but that good determinations (less than 0.1% error) can be made with a cell which restricts potential overshoot to less than 10 mv. The new cell design, which provides adequate potential control, permits high titration speeds, and completely contains radioactive contamination, is presented. (auth)
- Research Organization:
- General Electric Co., Richland, Wash. (USA). Hanford Atomic Products Operation
- DOE Contract Number:
- AT(45-1)-1350
- NSA Number:
- NSA-29-026583
- OSTI ID:
- 4330216
- Report Number(s):
- HW--80365
- Country of Publication:
- United States
- Language:
- English
Similar Records
APPLICATIONS OF CONTROLLED POTENTIAL COULOMETRY
Uranium by controlled-potential coulometry
THE DETERMINATION OF PLUTONIUM IN IRRADIATED URANIUM FUEL SOLUTIONS BY CONTROLLED POTENTIAL COULOMETRY
Technical Report
·
Fri May 25 00:00:00 EDT 1962
·
OSTI ID:4770305
Uranium by controlled-potential coulometry
Journal Article
·
Wed Dec 31 23:00:00 EST 1980
· Annu. Book ASTM Stand.; (United States)
·
OSTI ID:5680675
THE DETERMINATION OF PLUTONIUM IN IRRADIATED URANIUM FUEL SOLUTIONS BY CONTROLLED POTENTIAL COULOMETRY
Technical Report
·
Tue Dec 09 23:00:00 EST 1958
·
OSTI ID:4254039