Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

TWO-CRYSAAL-COMPTON-SPECTROMETER WITH IMPROVED GEOMETRICAL EFFICIENCY

Journal Article · · Nuclear Instr.

The low detection efficiency of the Compton spectrometer is disadvantageous in nuclear decay scheme work with low-level or short-life activities. Omitting the gamma-collimator and placing the source bstween the two crystals improves the efficiency. To reduce the number of gamma coincidences caused by priunary quanta in the two crystals, relative to the Comptonscattering coincidences, a kicksorter was applied which accepts only those pulses from crystal 2 whose energy falls in the narrow band of back-scattered Compton quanta. The use of anthacene as the scattering crysial is advisable if the poorer resolution of this scintillator can be tolerated. This new version of the Compton spectrometer gives an increase of over-all efficiency by about two orders of magnitude. (A.C.)

Research Organization:
Univ. of Vienna
NSA Number:
NSA-12-017888
OSTI ID:
4292410
Journal Information:
Nuclear Instr., Journal Name: Nuclear Instr. Vol. Vol: 3
Country of Publication:
Country unknown/Code not available
Language:
English

Similar Records

SCINTILLATION SPECTROMETER WITH SUMMATION OF PULSES FROM TWO CRYSTALS
Journal Article · Tue Dec 31 23:00:00 EST 1957 · Ceskoslov. casopis pro fysiku · OSTI ID:4144136

NaI(Tl) SUMMING COMPTON SPECTROMETER
Journal Article · Wed Nov 30 23:00:00 EST 1960 · Rev. Sci. Instr. · OSTI ID:4109735

A HIGH RESOLUTION COMPTON-SPECTROMETER USING A SEMI-CONDUCTOR DETECTOR
Journal Article · Sun Sep 01 00:00:00 EDT 1963 · Nucl. Instr. Methods · OSTI ID:4119524