METHODS OF FABRICATING ALPHA PARTICLE SCINTILLATION SCREENS

Technical Report ·
OSTI ID:4228236

A suitable scintillation screen for use in an alpha detection system was developed. Zinc sulfide was used as the scintillation phosphor and was deposited by means of a bending solution onto a transparent backing plate. To exclude ambient light and to afford some amount of light recoated 1/4 mil Mylar sheet. The Mylar sheet was bonded to the phosphor by means of an edging of 100% solids adhesive. Since the scintillation screen was to be covered bv a fine wire mean the finished detector (with a moderately rugged light-tight covering) was considered adequate. The over-all alpha particle energy attenuation was less than 1 Mev. (auth)

Research Organization:
Naval Radiological Defense Lab., San Francisco
NSA Number:
NSA-14-001649
OSTI ID:
4228236
Report Number(s):
USNRDL-TR-349
Country of Publication:
Country unknown/Code not available
Language:
English

Similar Records

A more rugged ZnS(Ag) alpha scintillation detector
Conference · Sat Mar 31 23:00:00 EST 1990 · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (USA) · OSTI ID:6869202

Thin film techniques for developing a more rugged alpha scintillation detector
Conference · Thu Dec 31 23:00:00 EST 1987 · OSTI ID:6694971

A ruggedized ZnS(Ag)/epoxy alpha scintillation detector
Conference · Sun Dec 31 23:00:00 EST 1989 · OSTI ID:6943879